ANY-maze Help > The ANY-maze reference > The Tests page > Running tests > General information about running tests > Testing do's and don'ts

Testing do's and don'ts

Introduction

To ensure ANY-maze testing works smoothly there are a few things it's good to be aware of. In fact most of them are either obvious or don't really require any change from what you would probably do anyhow, but it's still good to know what they are.

 •A general point - ANY-maze is always watching 
 •When you start ANY-maze (or when you switch the cameras on) try to make sure no one is in the camera's view 
 •Before a test starts there should be a period of at least 4 seconds during which the apparatus is "empty" and no one is in the camera's view 
 •Don't leave things behind in the camera's view when you start a test 
 •Once a test has started you CAN walk around in the area outside the apparatus 
 •If your apparatus has a transparent lid or cover of some kind then keep it closed ALL THE TIME except, obviously, when you're putting an animal in or taking an animal out of the apparatus. 

First a general point - ANY-maze is always watching!

You might think that ANY-maze is only analysing images while a test is running but this isn't actually the case. It's always analysing the video pictures it receives and it uses them to better understand your apparatus. This fact is partly responsible for some of the other recommendations given below.

When you start ANY-maze (or when you switch the cameras on) try to make sure no one is in the camera's view

This is not critical and usually just happens anyway because to start ANY-maze you'll need to be at the computer and, ideally, the computer won't be in the camera's view.

If there is someone visible when you start ANY-maze then, when the person LEAVES the camera's view, a 'person' icon will probably be shown in the top-left corner of the video picture (usually this icon is displayed when someone ENTERS the camera's view) - just click the icon and you'll fix the problem.

Before a test starts there should be a period of at least 4 seconds during which the apparatus is 'empty' and no one is in the camera's view

In order for ANY-maze to quickly identify the animal at the start of a test it should have had a chance to see what the empty apparatus looks like before the test starts (this is one of those things it's analysing when tests aren't running). Here 'empty' means not just that there's no animal in the apparatus but also that you are not anywhere in the video picture either.

In fact this is another thing which usually just happens anyway because you obviously have to get the animal before you start a test and this usually takes at least 4 seconds and while you're doing it you probably won't be visible to the camera.

Don't leave things behind in the camera's view when you start a test

Generally, at the start of a test you will walk into the camera's view, place an animal on the apparatus and then walk away again - as ANY-maze sees you do this it will start the test.

However, if you leave something behind in the camera's view (such as the animal's box) then ANY-maze will probably think there's still someone in the picture and it won't start the test.

Note: If you don't intend to use the auto-start feature then you don't need to worry about this.

Once a test has started you CAN walk around in the area outside the apparatus (although there are usually good scientific reasons for avoiding this)

If you reappear in the camera's view AFTER a test has started you won't interfere with the tracking. However, you should be careful not to come between the camera and the apparatus as this will almost certainly have an adverse effect.

If your apparatus has a transparent lid or cover of some kind then keep it closed ALL THE TIME except, obviously, when you're putting an animal in or taking an animal out of the apparatus

This is rarely an issue as most apparatus doesn't have a lid or cover; but if it does, and the camera has too look through it to see the animal, then it's important that it's always looking through it even when tests aren't running -  again this is related to the fact that ANY-maze is always analysing the apparatus and it needs to be looking at the apparatus as it will appear while tests are running, i.e. with any lid or cover closed.

© Copyright 2003-2026 Stoelting Co. All rights reserved