ANY-maze Help > I/O devices supported by ANY-maze > The ANY-maze Operant interfaces > The ANY-maze Operant interface L > The ANY-maze Operant interface L specification

The ANY-maze Operant interface L specification

Introduction

The ANY-maze Operant interface L is a drop-in replacement for a Lafayette Instrument ABET Mini I/O Module Connection Block.

Inputs

The interface has 8 inputs which are identical to the 8 inputs of the ABET Mini I/O Module Connection Block. These are used to connect components such as levers.

The physical connectors are the same screw terminal blocks used by Lafayette device, with the lower, front row of connectors being the inputs and the higher, rear row of connectors being Common (i.e. GND). Connecting an input to Common activates the input.

Outputs

The interface has 16 outputs which are identical to the 16 outputs of the ABET Mini I/O Module Connection Block. These are used to connect components such as pellet dispensers.

The physical connectors are the same screw terminal blocks used by Lafayette device, with the lower, front row of connectors being the outputs and the higher, rear row of connectors being 28V. When activated, an output will be switched to GND.

The interface also includes 5 Intensity outputs on telephone-style RJ12 connectors, these are identical to the Intensity connectors of the Lafayette device. Each intensity connector uses 3 of the output lines, with Intensity 1 using Outputs 1-3, Intensity 2 using Outputs 4-6, and so on. The intensity connectors are used by specific components and, in effect, just provide a convenient way to connect three outputs at once.

Aux outputs

The interface includes 2 Aux connectors. These are not present on the Lafayette ABET Mini I/O Module Connection Block but instead replicate Aux outputs found on the rear panel of some Lafayette interfaces to which the ABET Mini I/O Module Connection Blocks connect.

The Aux outputs are simple relays: when an Aux connector is activated the 2 pins of the relevant connector are connected together - like a switch. A typical use for the Aux connectors is to control a device such as the Lafayette Instrument HSCK100AP Shocker by connecting it to the shocker's Remote on/off jacks.

Test control push button

The red, test control push button, is mounted on the top of the interface. It can be configured to start a test when pushed - this is very helpful as it means you can quickly start a test after placing an animal in the cage.

DC power

The interface is power from a 28V DC power supply which should be connected to the 5.5mm, centre-positive, barrel connector next to the USB connector.

LED indicator

There is an LED indicator on the front of the interface which shows the interface's status:

Off No USB or 28V power connected.
Red28V Power supply is connected, but USB is not connected.
Brief flashing greenUSB is connected, but the 28V power supply is disconnected or switched off.
Solid greenBoth USB and power are connected and switched on - the interface is ready to use.
Even flashing greenThe interface has been set to identify itself.

USB connector

The interface connects to the computer via a USB 2 'B' connector. The interface will work correctly when connected to either a USB 2 or USB 3 connector on the computer.

Physical

The interface has the same footprint as the Lafayette ABET Mini I/O Module Connection Block and can be screwed into the same mounting holes in the base plate of a Lafayette operant cage.

© Copyright 2003-2026 Stoelting Co. All rights reserved