ANY-maze Help > I/O devices supported by ANY-maze > Ugo Basile instruments supported by ANY-maze > The Ugo Basile Fear Conditioning Cage > The Ugo Basile Fear Conditioning Cage external digital outputs The Ugo Basile Fear Conditioning Cage external digital outputs
IntroductionThere are four external TTL outputs available on an Ugo Basile Fear conditioning Cage. TTL is a standard for digital signals and many devices that accept digital inputs use TTL logic levels. For example, some pellet dispensers will dispense a pellet in response to a TTL 'pulse' on their input. The Fear conditioning cage TTL outputs specification
Connecting an output to somethingTo control a device that has a TTL level input, you simply need to connect an output (and ground) to the device's input. If you want to control something that doesn't have a TTL level input, for example, you want to switch a lamp on and off, then you will need to use some additional circuit which will be controlled by the TTL output. This might not be as complicated as it sounds; for example, a Google search for 'TTL Relay Board' will probably turn up a number of suitable and inexpensive devices. Using an output within an experimentIn order to use an output in an experiment, it must be included in the protocol. This is a simple process and is described in detail in the Setting up an on/off output topic. As well as including the output in the protocol, you will probably also want to include one or more procedures. For example, you could use a procedure to determine when the animal freezes for a certain duration and then turn the output on. See also:
© Copyright 2003-2026 Stoelting Co. All rights reserved ANY-maze help topic T1261 |