Setting a temperature controller to ramp to a different temperature during the test

The ramp settings described in this topic are not available for the OPAD temperature controller - they are replaced by the OPAD temperature cycle settings.

In brief

You can use Ramp fields on the temperature controller's settings page to specify that the temperature of the controller should ramp from the start temperature to some other temperature over the programmed duration of the test. The temperature will start ramping as soon as the test begins.

Details

If you select this option, then the temperature controller will 'ramp' from whatever temperature it has at the start of the test (which you can specify using the start temperature field) to the value you specify. The ramp will be calculated so that the temperature at the programmed end of the test will be the value you specify, and the ramp will be linear. For example, if you specify that the start temperature is 20°C and the end temperature is 30°C, and the test is programmed to last 100 seconds, then the temperature will change 0.1°C each second.

There are a few things to be aware of:

 •The test may not last for the programmed duration; for example, it might be ended by an procedure before this - in which case, the temperature will never reach the value you set. 
 •The device may not be able to reach the temperature you specify, in which case it will just get as hot or cold as it can. 
 •While the temperature will in theory ramp in a linear way, the device will actually 'step' the temperature. The size of the steps will depend on the device, so for example in a device which can only set the temperature to a resolution of 1°C, then the steps will be 1°C and so, in the case of ramping from 20°C to 30°C during a 100 second test, the device will actually change temperature by 1°C every 10 seconds. 